StatementOfDeficiencies.comSurvey findings and how to respond Book a free consultation
Home › PT/INR

Prothrombin Time (PT/INR): Survey Deficiencies and CLIA Citations

Most PT/INR citations come from five places: quality control (42 CFR 493.1256), calibration verification (42 CFR 493.1255), proficiency testing or accuracy verification, method verification (42 CFR 493.1253) and test reporting, including critical values (42 CFR 493.1291). Keep the records for each area current and you remove most of the risk.

Last reviewed Nov 24, 2026 · By the Sound Lab Consultants regulatory team · Sources listed below

Where PT/INR citations come from

Survey findings about PT/INR almost always trace to a short list of 42 CFR 493 standards. This table shows each one and the records a surveyor will usually ask to see.

AreaRegulationRecords surveyors ask for
QC42 CFR 493.1256, 42 CFR 493.1269Daily PT/INR QC records, Levey-Jennings charts, corrective action log, control lot ranges
Calibration verification42 CFR 493.1255Six-month PT/INR calibration verification and reportable range data with director review
Proficiency testing42 CFR 493.801, Subpart HPT/INR PT enrollment, attestation statements, graded reports, investigation of every miss
Method verification42 CFR 493.1253PT/INR verification study: accuracy, precision, reportable range, reference interval check, director sign-off
Test report and critical values42 CFR 493.1291Critical value list, call log with read-back, report format showing PT/INR reference intervals
PersonnelSubpart MTraining and competency records for everyone who runs PT/INR

PT status for PT/INR

Prothrombin Time (PT/INR) is a regulated analyte (hematology (coagulation), 42 CFR 493.941). CLIA PT acceptance limit: target ±15%.

Specimen problems that turn into PT/INR findings

Preanalytical errors become citations when the lab has no written rule for them, or does not follow its own rule.

If PT/INR was cited on your survey

  1. Identify the exact standard cited and the records the surveyor reviewed.
  2. Look back at patient results that could have been affected and decide whether corrected reports are needed.
  3. Fix the system, not just the instance, and assign monitoring.
  4. Write the plan of correction. See the CLIA plan of correction guide.

Related tests

Full guide: Prothrombin Time (PT/INR) testing in the clinical lab: CLIA, PT and QC requirements on soundlabconsult.com, part of the Laboratory Test Compliance Library.

For laboratory compliance and operations only, not medical advice or patient care guidance.

Facing a statement of deficiencies?

Sound Lab Consultants helps clinical laboratories respond to statements of deficiencies, from root cause to plan of correction to revisit.

Book a free consultation

Related guides on soundlabconsult.com:

Frequently asked questions

What are the most common PT/INR survey deficiencies?

Missing or undocumented QC, missed calibration verification, PT or accuracy verification gaps, no method verification study, and undocumented critical value calls for PT/INR. These are general categories, not statistics.

Which regulation covers PT/INR quality control?

42 CFR 493.1256 sets the control procedure requirements, with coagulation rules in 42 CFR 493.1269.

Is PT/INR on the CLIA PT analyte list?

Prothrombin Time (PT/INR) is a regulated analyte (hematology (coagulation), 42 CFR 493.941). CLIA PT acceptance limit: target ±15%.

Sources

More on StatementOfDeficiencies.com