Chlamydia and Gonorrhea NAAT: Survey Deficiencies and CLIA Citations
Most CT/NG NAAT citations come from five places: quality control (42 CFR 493.1256), calibration verification (42 CFR 493.1255), proficiency testing or accuracy verification, method verification (42 CFR 493.1253) and test reporting, including critical values (42 CFR 493.1291). Keep the records for each area current and you remove most of the risk.
Where CT/NG NAAT citations come from
Survey findings about CT/NG NAAT almost always trace to a short list of 42 CFR 493 standards. This table shows each one and the records a surveyor will usually ask to see.
| Area | Regulation | Records surveyors ask for |
|---|---|---|
| QC | 42 CFR 493.1256 | Daily CT/NG NAAT QC records, Levey-Jennings charts, corrective action log, control lot ranges |
| Proficiency testing | 42 CFR 493.801, Subpart H | CT/NG NAAT PT enrollment, attestation statements, graded reports, investigation of every miss |
| Method verification | 42 CFR 493.1253 | CT/NG NAAT verification study: accuracy, precision, reportable range, reference interval check, director sign-off |
| Test report and critical values | 42 CFR 493.1291 | Critical value list, call log with read-back, report format showing CT/NG NAAT reference intervals |
| Personnel | Subpart M | Training and competency records for everyone who runs CT/NG NAAT |
PT status for CT/NG NAAT
No analyte-level limit; nonwaived labs enroll in bacteriology PT for the services they perform (42 CFR 493.911).
Specimen problems that turn into CT/NG NAAT findings
Preanalytical errors become citations when the lab has no written rule for them, or does not follow its own rule.
- Use only the collection kits and specimen types the assay is cleared for.
- Amplicon contamination control: separate work areas, cleaning and workflow are essential for NAAT.
- Public health reporting requirements apply to positive results.
If CT/NG NAAT was cited on your survey
- Identify the exact standard cited and the records the surveyor reviewed.
- Look back at patient results that could have been affected and decide whether corrected reports are needed.
- Fix the system, not just the instance, and assign monitoring.
- Write the plan of correction. See the CLIA plan of correction guide.
Related tests
- Trichomonas vaginalis Testing (full guide)
- HIV Antibody (Anti-HIV)
- Urinalysis (Dipstick and Microscopic) (full guide)
Full guide: Chlamydia and Gonorrhea NAAT testing in the clinical lab: CLIA, PT and QC requirements on soundlabconsult.com, part of the Laboratory Test Compliance Library.
Facing a statement of deficiencies?
Sound Lab Consultants helps clinical laboratories respond to statements of deficiencies, from root cause to plan of correction to revisit.
Related guides on soundlabconsult.com:
Frequently asked questions
What are the most common CT/NG NAAT survey deficiencies?
Missing or undocumented QC, missed calibration verification, PT or accuracy verification gaps, no method verification study, and undocumented critical value calls for CT/NG NAAT. These are general categories, not statistics.
Which regulation covers CT/NG NAAT quality control?
42 CFR 493.1256 sets the control procedure requirements.
Is CT/NG NAAT on the CLIA PT analyte list?
No analyte-level limit; nonwaived labs enroll in bacteriology PT for the services they perform (42 CFR 493.911).
Sources
- 42 CFR 493.911, Bacteriology
- 42 CFR 493.1256, Control procedures
- 42 CFR 493.1255, Calibration and calibration verification
- 42 CFR 493.1253, Establishment and verification of performance specifications
- 42 CFR 493.1291, Test report
- 42 CFR 493.1236, Evaluation of proficiency testing performance
- CMS CLIA Year 2024 Top 10 Standard- and Condition-Level Deficiencies
- CMS State Operations Manual, Appendix C (Laboratories)