AST (Aspartate Aminotransferase): Survey Deficiencies and CLIA Citations
Most AST citations come from five places: quality control (42 CFR 493.1256), calibration verification (42 CFR 493.1255), proficiency testing or accuracy verification, method verification (42 CFR 493.1253) and test reporting, including critical values (42 CFR 493.1291). Keep the records for each area current and you remove most of the risk.
Where AST citations come from
Survey findings about AST almost always trace to a short list of 42 CFR 493 standards. This table shows each one and the records a surveyor will usually ask to see.
| Area | Regulation | Records surveyors ask for |
|---|---|---|
| QC | 42 CFR 493.1256 | Daily AST QC records, Levey-Jennings charts, corrective action log, control lot ranges |
| Calibration verification | 42 CFR 493.1255 | Six-month AST calibration verification and reportable range data with director review |
| Proficiency testing | 42 CFR 493.801, Subpart H | AST PT enrollment, attestation statements, graded reports, investigation of every miss |
| Method verification | 42 CFR 493.1253 | AST verification study: accuracy, precision, reportable range, reference interval check, director sign-off |
| Test report and critical values | 42 CFR 493.1291 | Critical value list, call log with read-back, report format showing AST reference intervals |
| Personnel | Subpart M | Training and competency records for everyone who runs AST |
PT status for AST
AST (Aspartate Aminotransferase) is a regulated analyte (routine chemistry, 42 CFR 493.931). CLIA PT acceptance limit: target value ±15% or ±6 U/L (greater).
Specimen problems that turn into AST findings
Preanalytical errors become citations when the lab has no written rule for them, or does not follow its own rule.
- Hemolysis (in vitro) during collection or transport: red cells contain much more AST than plasma, so even slight hemolysis raises results.
- Delayed separation can raise AST; centrifuge within the time your procedure allows.
If AST was cited on your survey
- Identify the exact standard cited and the records the surveyor reviewed.
- Look back at patient results that could have been affected and decide whether corrected reports are needed.
- Fix the system, not just the instance, and assign monitoring.
- Write the plan of correction. See the CLIA plan of correction guide.
Related tests
- ALT (Alanine Aminotransferase) (full guide)
- LDH (Lactate Dehydrogenase) (full guide)
- Creatine Kinase (CK) (full guide)
- Bilirubin, Total (full guide)
Full guide: AST (Aspartate Aminotransferase) testing in the clinical lab: CLIA, PT and QC requirements on soundlabconsult.com, part of the Laboratory Test Compliance Library.
Facing a statement of deficiencies?
Sound Lab Consultants helps clinical laboratories respond to statements of deficiencies, from root cause to plan of correction to revisit.
Related guides on soundlabconsult.com:
Frequently asked questions
What are the most common AST survey deficiencies?
Missing or undocumented QC, missed calibration verification, PT or accuracy verification gaps, no method verification study, and undocumented critical value calls for AST. These are general categories, not statistics.
Which regulation covers AST quality control?
42 CFR 493.1256 sets the control procedure requirements.
Is AST on the CLIA PT analyte list?
AST (Aspartate Aminotransferase) is a regulated analyte (routine chemistry, 42 CFR 493.931). CLIA PT acceptance limit: target value ±15% or ±6 U/L (greater).
Sources
- 42 CFR 493.931, Routine chemistry
- 42 CFR 493.1256, Control procedures
- 42 CFR 493.1255, Calibration and calibration verification
- 42 CFR 493.1253, Establishment and verification of performance specifications
- 42 CFR 493.1291, Test report
- 42 CFR 493.1236, Evaluation of proficiency testing performance
- CMS CLIA Year 2024 Top 10 Standard- and Condition-Level Deficiencies
- CMS State Operations Manual, Appendix C (Laboratories)