Alpha-Fetoprotein (AFP), Tumor Marker: Survey Deficiencies and CLIA Citations
Most AFP Tumor Marker citations come from five places: quality control (42 CFR 493.1256), calibration verification (42 CFR 493.1255), proficiency testing or accuracy verification, method verification (42 CFR 493.1253) and test reporting, including critical values (42 CFR 493.1291). Keep the records for each area current and you remove most of the risk.
Where AFP Tumor Marker citations come from
Survey findings about AFP Tumor Marker almost always trace to a short list of 42 CFR 493 standards. This table shows each one and the records a surveyor will usually ask to see.
| Area | Regulation | Records surveyors ask for |
|---|---|---|
| QC | 42 CFR 493.1256 | Daily AFP Tumor Marker QC records, Levey-Jennings charts, corrective action log, control lot ranges |
| Calibration verification | 42 CFR 493.1255 | Six-month AFP Tumor Marker calibration verification and reportable range data with director review |
| Proficiency testing | 42 CFR 493.801, Subpart H | AFP Tumor Marker PT enrollment, attestation statements, graded reports, investigation of every miss |
| Method verification | 42 CFR 493.1253 | AFP Tumor Marker verification study: accuracy, precision, reportable range, reference interval check, director sign-off |
| Test report and critical values | 42 CFR 493.1291 | Critical value list, call log with read-back, report format showing AFP Tumor Marker reference intervals |
| Personnel | Subpart M | Training and competency records for everyone who runs AFP Tumor Marker |
PT status for AFP Tumor Marker
Alpha-Fetoprotein (AFP), Tumor Marker is a regulated analyte (general immunology, 42 CFR 493.927). CLIA PT acceptance limit: target value ± 20%.
Specimen problems that turn into AFP Tumor Marker findings
Preanalytical errors become citations when the lab has no written rule for them, or does not follow its own rule.
- AFP as a tumor marker and AFP for prenatal screening are different uses; the PT listing is for the tumor marker.
- Different AFP assays are not interchangeable for serial monitoring; tell clients when you change methods.
If AFP Tumor Marker was cited on your survey
- Identify the exact standard cited and the records the surveyor reviewed.
- Look back at patient results that could have been affected and decide whether corrected reports are needed.
- Fix the system, not just the instance, and assign monitoring.
- Write the plan of correction. See the CLIA plan of correction guide.
Related tests
- CEA (Carcinoembryonic Antigen) (full guide)
- CA 125 (Cancer Antigen 125) (full guide)
- hCG (Human Chorionic Gonadotropin) (full guide)
Full guide: Alpha-Fetoprotein (AFP), Tumor Marker testing in the clinical lab: CLIA, PT and QC requirements on soundlabconsult.com, part of the Laboratory Test Compliance Library.
Facing a statement of deficiencies?
Sound Lab Consultants helps clinical laboratories respond to statements of deficiencies, from root cause to plan of correction to revisit.
Related guides on soundlabconsult.com:
Frequently asked questions
What are the most common AFP Tumor Marker survey deficiencies?
Missing or undocumented QC, missed calibration verification, PT or accuracy verification gaps, no method verification study, and undocumented critical value calls for AFP Tumor Marker. These are general categories, not statistics.
Which regulation covers AFP Tumor Marker quality control?
42 CFR 493.1256 sets the control procedure requirements.
Is AFP Tumor Marker on the CLIA PT analyte list?
Alpha-Fetoprotein (AFP), Tumor Marker is a regulated analyte (general immunology, 42 CFR 493.927). CLIA PT acceptance limit: target value ± 20%.
Sources
- 42 CFR 493.927, General immunology
- 42 CFR 493.1256, Control procedures
- 42 CFR 493.1255, Calibration and calibration verification
- 42 CFR 493.1253, Establishment and verification of performance specifications
- 42 CFR 493.1291, Test report
- 42 CFR 493.1236, Evaluation of proficiency testing performance
- CMS CLIA Year 2024 Top 10 Standard- and Condition-Level Deficiencies
- CMS State Operations Manual, Appendix C (Laboratories)